Summary
Hardware, Detached Light Pipe Version:
DUT History
June 2025 Lee has received a VentMon during a visit to Austin TX.
I think it is version: 0.5
There is no serial number on this device.
Details
What I think would be called a light pipe has become detached and I have secured it with transparent tape.
Boot up message
14:15:59.554 -> OLED begun
14:16:06.635 -> Starting VentMon...Checking sensors...Ambient Pressure Sensor : NO
14:16:07.845 -> Inspiratory Pressure Sensor : BME280
14:16:07.937 -> Found O2 Sensor: NO
14:16:08.903 -> sensor serial number: 862B8B9F
14:16:08.903 -> sensor article number: 118ABD01
14:16:08.903 ->
14:16:08.903 -> read scale factor: 120.00
14:16:08.903 -> read flow offset: 32768.00
14:16:08.903 -> flow : 0.00 slm
14:16:09.034 -> flow : -0.03 slm
14:16:09.126 -> flow : 0.00 slm
14:16:09.218 -> flow : 0.00 slm
14:16:09.302 -> flow : 0.00 slm
14:16:09.431 -> Found Differential Pressure Sensor: YES
14:16:09.431 -> ----------------- SENSOR TEST COMPLETE -----------------
14:16:09.431 -> Wifi not an AP
14:16:09.431 -> Config via serial...WiFi DISABLED
14:16:09.934 -> Enter 'w' to enable WiFi, 'x' to disable WiFi
14:16:09.934 -> Enter 'c' to re-enter this configuration while running.
14:16:09.968 -> { "event": "M", "type": "P", "ms": 10848, "loc": "I", "num": 0, "val": 10085 }
14:16:09.968 -> { "event": "M", "type": "T", "ms": 10850, "loc": "I", "num": 0, "val": 2545 }